Test Items:
|
22 parameters: RBC, HGB, WBC, LY#,
MO#, GR#, LY%, MO%, GR%, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PCT,
PDW |
|
| Histogram Display: |
White Blood Cell Histogram
(Colour) |
| Red Blood Cell Histogram (Colour) |
| Platelet Histogram (Colour) |
|
Linearity Range:
|
WBC (×109/L): 0.0-99.9 |
| RBC (×1012/L): 0.00-99.9 |
| HGB (g/L): 20-200 |
| MCV (fL): 40.0-150 |
| PLT(×109/L): 0.0-999 |
|
Precision (CV):
|
WBC (×109/L) < 2%
(5.0-10.0×109/L) |
| RBC (×1012/L) <1.5%
(3.5-6.0×1012/L) |
| HGB (g/L) <1.5%
(110-180g/L) |
| MCV (fL) <0.5%
(80.0-110fL) |
| PLT (×109/L) <4.0%
(100-500×109/L) |
|
Data Storage:
|
25,000 results |
|
Dilution Ratio:
|
WBC / HGB 1:500 |
| RBC / PLT 1:50000 |
|
Sample volume:
|
Periphery/whole blood: 20µl |
|
Display:
|
640×480 pixels, 9.4" TFT |
|
Printer:
|
Built-in thermal printer, paper width:
5.7cm |
|
Test Time:
|
Less than 40 seconds |
|
Continuous Work Time:
|
More than 8 hours |
|
Power:
|
220V±22V 50Hz±1Hz |
|
Fuse:
|
BLX-1 50×20-2A |
|
Input Power:
|
≤180VA |
|
Interfaces:
|
RS232
joint |
|
|
|
|
|
|